Radiation Concepts:

Scatter/ Wedges / Equivalent Square/ Field matching/ Beam Blocking

In EBRT, the Radiation beam exits from the linac head in form of high energy photons.

Isocentre: The isocentre is the point in space about which the gantry of the linear accelerator, the treatment head of the linear accelerator, and the couch rotate.

Beam profile: Transverse dose measurements are performed in a plane perpendicular in direction to the radiation beam, and at a given depth in the phantom. These are known dose profiles. Beam profile is designed to be FLAT AT DEPTH of 10cm typically (dog ears or M shaped at surface)

SCATTER

Scatter: Scatter Radiation is a type of secondary radiation that occurs when the primary radiation beam intercepts any object, causing some x-rays to be scattered. Main components of scatter consists of : Head scatter (Flattening Filter and collimator scatter) and phantom scatter (in patient) 

Scatter factor: total dose/primary dose (dose without scatter) 

Peak Scatter Factor: SF at d max 

Back Scatter Factor: SF at surface 

  1. <45kV BSF 1.01  (PE effect predominates, electrons not scattered back with sufficient energy to penetrate medium) 

2. 300kV  BSF upto 1.5 (compton predominates, back scattered electrons have sufficient energy to reach surface).

3. > 2MV BSF 1.02 (mainly forward scatter due to high energy of particles) 

ELECTRON contamination: electrons produced in head of linac that contaminate dose at surface. Higher with increased use of trays, wedges, increased field size.

EQUIVALENT SIZE of fields: 

For a rectangle field: Eq. square = 2axb / a+b 

But for a field of any shape (rectangle with a corner cut out) = 4 x area / perimeter  

Irregular field shapes/blocking. Same Radiation field area, but different shapes = different dose

Dose from from primary RT same 

Dose from secondary radiation in reduced. Dose in centre of field depends on scatter from edges of field 

THEREFORE DOSE DEPENDS ON AREA and SHAPE. 

Primary dose calculated depending on linac head geometry 

Scatter contribution (head scatter) calculated by integrating Scatter air ratios for shape. (divide shapes into segments of different radii from a point centre and add them up) 

Hence sector integration or Clarkson method to calculate dose. 

WEDGES:

  1. Minimise dose gradient in treatment volume 

2. Minimise hot spots 

3. Compensate for patient shape, volume shape 

IEC definition of Wedge angle: angle of the isododose curve to a plane at right angles to the central axis at a depth of 10cm 

A wedge can be physical or lump of lead – or can be MLC 

Upto 60 degrees normally 

COPLANAR vs NON COPLANAR 

Coplanar – irradiated volume is all in the same ‘slab’ of tissue. Sup and Inf borders can easily be defined. All beam axes in the same plane 

VMAT can be co-planar (traditional) or non co-planar 

FIELD MATCHING

Two abutting fields from the same direction – matched so that 50% isodose at a certain depth overlap. Accept shallower depths will have less dose, and deeper will have greater (more field overlap) 

If the gantry angles are rotated or beams are half beam blocked such that the 50% idosose lines are parallel, matched at all depths (remove beam divergence at match line) . See this brilliant video by SadieRayDiation on youtube.

BEAM BLOCKING

To minimise divergence from tangential opposing pair of beams – can use half beam block. (isocentre at mid point of one edge of beam) 

Can use 3/4th beam block also for breast treatment such that superior and posterior edges of beam are straight or non divergent. Isocentre in then in a corner of the field.  

Leave a Reply

Fill in your details below or click an icon to log in:

WordPress.com Logo

You are commenting using your WordPress.com account. Log Out /  Change )

Google photo

You are commenting using your Google account. Log Out /  Change )

Twitter picture

You are commenting using your Twitter account. Log Out /  Change )

Facebook photo

You are commenting using your Facebook account. Log Out /  Change )

Connecting to %s