IMAGING DURING TREATMENT
CT – linear attenuation coefficient for each pixel is compared to water and gives us Hounsefield number
CT numbers represent Electron density of material
Compton attenuation is proportional to Electron density
MRI better for soft tissue contrast. But no electron density information. Prone to geometric uncertainty.
PET : F18 is a positron emitter
F18 FDG cannot be distinguished by the body from glucose
positron annihilated by an electron and 2 photons in opposite direction produced. Detecting these photons helps understand where the original position, i.e. glucose uptake was
Patient Set up for External beam RT:
tattoos and fixed room lasers
Immobilisation devices
treatment field – corresponds to a light field from treatment head and centre seen with cross wires
checking SSD (source to surface distance)
imaging protocols or cone beam CT verification
Immobilisation considerations
Reproducibility
Comfort
Limited normal tissues radiated (wing board in lung to protect arms)
Technique and allowing gantry rotation freely
Verification imaging
Anatomy comparison between planning and treatment times
2D MV or kV images can be acquired during each pretreatment episode
2D verification images for comparison can be a DRR (digitally reconstructed radiograph) from planning CT
3D CBCT (Cone beam CT)
Captured with a single rotation of X ray tube and detector while patient is on the treatment couch.
Detector is a single square panel (as opposed to diagnostic CT which is a thin linear ray of detectors and reads helically through patient slice by slice)
Increased noise of CBCT due to detector geometry
CBCT has good bony anatomy and some soft tissue information
good resolution in all axes
Patient breathing motion correction
Decrease motion – abdominal compression.
Active breathing coordinator, (block nose, breath into a tube which is attached to a machine through mouth. Has a resistance that stops you from breathing out and allows treatment to be delivered during the time that fixed volume of air in lung – inspiration and breast moved away from lung) or gating (using a marker on the chest and as it moves can only treat when the marker is at a certain range – it will move with chest wall)
What is a 4D CT
Low pitched CT scan (slow movement through scanner)
Images each slice of the patient through all phases of respiratory cycle
Essentially a series of respiratory phase correlated 3DCTs
Respiratory waveforms are identified and split into bins and images grouped accordingly to make a series of 3DCT scans
CT scanner identifies phase of respiration either by an external marker that moves with chest wall, or implanted marker in tumour
MIP – max intensity of image at each point of breathing cycle combined
used for contouring
AVG is time averaged density of tumour- used for planning calculation
Min IP = shows location where the tumour is ALWAYS present. Might be Good for negative contrast contouring (liver tumours). NOT USED currently.
4DCBCT also rotates only once around patient – but very slowly! to capture all phases of respiratory cycle. Can take 5 min to take one scan. Uses often an internal marker to assess phase of respiration (diaphragm vs lung interface)
QUALITY ASSURANCE IN RT
Every machine and patient failure significant events go to CQC (Care Quality Commission). Machine errors also goes to HSE (Health and Safety Executive)
Significant Overexposure – 10% overexposure of dose for whole treatment or 20% for one fraction
Quality assurance vs Quality control
Assurance: Is a system in place or a list of instructions that must be followed to assure maintain quality
Quality control – is a measurement/means to check quality of a system
Quality control for LINACS
- Radiation output – checked daily for linac output to ensure the monitor units are constant as locally defined
2. Beam ENERGY
Determined by electron gun current and power of microwaves in wave guide
Measure of beam quality and how penetrative the beam is = TPR 20/10
TPR 20/10 : Tolerance is 1%
3. FLATNESS AND SYMMETRY
Beam profile affected by gun current, wave guide acceleration, beam steering
Tolerance 2% (feedback system so slightly more lax tolerance)
4. FIELD SIZE
Corresponding Light field from linac head – checked every morning. Easy to measure with graph paper – can just see the field.
X ray field size measurement – using MV image panel – EPID (Electronic portal imaging device)
5. Mechanical distance checks
Isocentre checked with cross wires
Source to surface distance checked with Optical distance indicator
6. IMAGING TESTS
Check imaging system for quality of image
Imaging data influences accuracy of treatments – geometric and dosimetric both
ELECTRO GUN influences: Beam Energy, symmetry and flatness
RF wave guide influences: Beam Energy, symmetry and flatness
STEERING SYSTEM influences: symmetry, flatness, leakage
Checks prior to treatment plan delivery
- Patient check, prescription check
2. Isocentre check
3. Machine parameters – is it physically deliverable
4. Optimisation check ( DVH)
5. MONITORING UNIT CHECK (different from output check!)
MU calculated by TPS should always be independently checked
(tolerance of about 3% as physical checks for simple plans assume all patient is water. National standards are 3% but for iMRT often dosimetric checks or independent computer verification can be used to check MUs and often achieves a much tighter tolerance)
Pre-treatment plan verification for IMRT
Can’t do is physically by physicists as too complex!
Physicists re-create a similar mock-plan in a phantom and measure dose with ion chamber or Thermoluminescent diode.
2D Film measurements
Radiographic or gafcrhomic films and measure dose exposure to film
Arrays of small diodes or ionisation chambers in an array that can be 2D or 3D
Initially phantom dose measurements done for every patient plan – but as more confident with Treatment Planning System, can decrease these
RECORD AND VERIFY SYSTEM
Data can be transferred from Treatment Planning Systems to treatment machines electronically
Records treatment data delivered automatically
GEOMETRIC ERROR SOURCES
1. Patient position :
Set up machine parameters corresponding to correct patient position
Couch angles, rotation etc can all be recorded and auto set
Treatment cannot be delivered if machine positions are not within set tolerances of treatment plan
Patient position verification
Upto 2mm tolerance
Tighter for SRS or SABR
2. internal motion
3. imperfect immobilisation
4. laser/machine instrument error
5. Human error
Can be systematic ( measurement device error, wrong immobilisation mask)
or random (patient movement)
Pre treatment imaging protocol – ONLINE IMAGING PROTOCOL while patient on bed
Can correct for shifts with CBCT prior to each fraction
OFFLINE IMAGING PROTOCOL
independent check of images taken after treatment completed
Set up checked frequently (every few fractions) and look for consistent shifts – then can correct for these
CBCT
Good contrast, wide field of view, 3D information, low dose
Poor resolution, takes time 2 min and patient might change position between CBCT and treatment, cannot show you treatment beam (only contours and patient position)
2D kV imaging
Same machinery as CBCT but only 2D images at 2 orthogonal angles
Quick, good bony anatomy, large field of view, good contrast
Often instead a DRR can be constructed from CBCT
2D MV imaging
Treatment head and EPID
Poor contrast (Compton predominant)
No extra dose is treatment dose is used to image a patient
Can see treatment field
IN VIVO DOSIMETRY
Diodes or TLD
Very rough estimate
Generally performed as check for gross errors at first dose
Not generally even used now as pretreatment verification is very good.
Can identify gross MU calculation errors, transfer of information from TPS to machine errors, gross patient set up error
Diodes positioned on patients skin during treatment
3-5% accuracy
Delivered dose behind diode is less (shadowing)
TLDs
rarely used as a lot of time for calibration of processing
small size – can be used to measure dose to small areas such as lens of eyes (over eyes)
Transit Dosimetry
Use exit dose to EPID (Electronic Portal Imaging Device) as a measure of dose delivered to patient. Back project exit dose into what would be expected in the middle of the patient. Very difficult in a practical sense.